8-8
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
2. Select [MONITOR SETUP]
3. Select [ALARM SETUP]
4. Select [ALM PAUSE TIME] to choose the alarm pause time from 1minute to X. X is the maximum
alarm pause time set in [MAX ALM PAUSE TIME].
8.7 Switching off Alarms
8.7.1
Switching off Individual Parameter Alarm
How to switch off individual parameter alarm, here we take ECG as an example:
1
Select ECG parameter area;
2
Select [ALM] to toggle between [ON] and [OFF].
When the alarm function of an individual parameter
corresponding parameter area.
is deactivated, the icon
will
appear in
8.7.2
Switching off All Parameter Alarms
Switching off all parameter alarms is password protected and can only be operated by the authorized
personnel. How to switch off all parameter alarms:
1. Access [MAIN MENU], select [MAINTAIN] and enter the password;
2. Select [ALARM SETUP];
3. Select [ALM] and toggle between [ON] and [OFF].
Moreover, the operator can be authorized to switch off all parameter alarms following the steps below:
1. Access [MAIN MENU], select [MAINTAIN] and enter the password;
2. Select [ALARM SETUP];
3. Select [ALM OFF Authorize];
4. Select [ON], and the operator can access [MAIN MENU] →[MONITOR SETUP] →[ALARM SETUP]
→[ALARM OFF] and select [ON] to switching off all parameter alarms.
When all parameter alarms are switched off, the system will give reminder tone periodically as the interval
set in [REMINDER INTERVAL]. During alarms off status:
1. For physiological alarms, no alarm lamp flashes, no alarms are sounded and no measured parameter
value and alarm limit flash. Physiological alarm message are not be displayed.
2. Prompt message “***ALARM OFF” is displayed in the physiological alarm message area with red
background.
3. For technical alarms, no alarms are sounded but with visual alarm and alarm message unchanged.
Содержание FM5000C
Страница 75: ...14 3 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure14 2 FHR Transducer ...
Страница 83: ...14 10 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual patient to avoid cross infection ...
Страница 97: ...15 13 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual baseline ...
Страница 102: ...16 1 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure 16 2 Positions of electrodes of 5 lead ECG cable ...
Страница 148: ...IV 2 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual PR alarm limit Lower limit 45 70 110 ...
Страница 155: ...Product Introduction Rear View Handle Paper tray Battery compartment Alarm Speaker Fetal Heart Speaker Nameplate ...