Cleaning stage 5:
MAIN WASH 2
Parameter [r45] = 0
‘Cleaning stage 5’ disabled
Parameter [r45] = 1
‘Cleaning stage 5’ execute one time
In case of cleaning method 'green' – it is set in programme step [n30] how often
the feed container is filled with water up to level 2 and water is pumped backup
to level 1. The drain valve remains closed up to cleaning step [n 5/n 6].
5
5
Settings for version 6.1.4 (
)
dosing in pre-hopper
P
5
arameter [r20 = 0] and parameter [n 0 = min. 2] since the detergent is added
when water is taken from the pre-hopper for the second time.
Water is taken in.
The parameter [n51] is used to stipulate whether water will be taken by level or
time.
If time-controlled, the water intake time is defined in parameter [n52].
If, in the event of intake by level, the level is not reached after a pre-set time
[r32], the cleaning cycle is interrupted. See section .5.3 - F44
7
The cleaning parameter [r54] is used to stipulate whether cold, hot or mixed
water is to be taken in.
Agitator and pump start to run. Water circulates in the tank.
Another option is to select an 'additional water intake time' using [n53].
Depending on the setting of the parameter [r20], acidic detergent is injected at
the end of the water intake time [n52] or the subsequent water intake time
[n53]. The duration of the detergent dosing is set in parameter [n102].
If a heater is present [r35], this is switched on. When the heating temperature
[n107] is reached, the heater is switched off again.
If the temperature is not reached within the pre-set time [r37], the control unit
continues with the program, but a fault report is displayed at the end. See sec-
tion 5.3 - F44
7.
If a switch-on delay [r36] has been parameterized for heating, heating starts
only after this time has elapsed. (This is to ensure that water is in the heater.)
Main circulation time: The water circulates for the pre-set time [n54]. During the
main circulating time, a post-heating temperature [n110] can be defined.
An additional time can be defined in parameter [n55], in which the drain valve
opens (agitator and cleaning pump continue to run)
Agitator and pump switch off.
Drain valve opens for the pre-set time [n56]. See also [n55].
Cleaning stage 6
: INTERMEDIATE RINSE 2
Parameter [r46] = 0
‘Cleaning stage 6’ disabled
Parameter [r46] = 1..5
‘Cleaning stage 6’ execute x-times
In case of cleaning method 'green' – it is set in programme step [n 0] how often
6
the feed container is filled with water up to level 2 and water is pumped back up
to level 1. As long as the rinsing pump is running, the drain valve is open.
Water is taken in.
The parameter [n61] is used to stipulate whether water will be taken by level or
time.
If time-controlled, the water intake time is defined in parameter [n62].
The cleaning parameter [r55] is used to stipulate whether cold, hot or mixed
water is to be taken in.
If, in the event of intake by level, the level is not reached after a pre-set time
[r32], the cleaning cycle is interrupted. See section .5.3 - F44
7
In case of green
n53 have to be
parameterized on „0“
In case of green
n51
n52
and
have to be
parameterized on „0“
In case of green
n61
n62
and
have to be
parameterized on „0“
6.3 Explanation and programming of the cleaning process
6. Cleaning
Page 25