1.3.2 Physical Specifications
Physical
Anodized aluminum enclosure.
GL (EyeLink 1000 Plus
Camera)
Standard thread (
¼
”-20) centered on optical axis from 3 sides.
M8 thread on front for DM and AM mounts.
Power requirements: +12VDC, 800 mA for camera alone, 1.8A maximum
when used with illuminator(s).
Imaging rates up to 2000 fps.
OC (Fiber Optic Camera
Head)
M3 tapped holes on body.
Adapter for
¼
”-20 mounting available.
Supply: 3.6-5.7VDC, 700mA via LEMO connector and power harness.
Imaging rates up to 2000 fps.
Non-magnetic (< 0.5g of iron or nickel).
Radio-quiet case and cable design for use in research environments.
FL-890, FL-940
(890/940 nm Fresnel
Illuminator Module)
Standard thread (
¼
”-20) at center and sides.
Adjustable focus and beamwidth.
Wavelength: 890 or 940 nm.
Supply: 3.6-5.7VDC, 1.1A via LEMO connector and power harness.
Non-magnetic (< 0.5g of iron or nickel).
Radio-quiet case and cable design for use in research environments.
PM-910, PM-940
(910/940 nm
Illuminator Module)
Wavelength: 910 or 940 nm.
Eye illumination level: less than 1 mW/cm
²
at >200mm from illuminator.
Illuminators powered from camera through supplied cables.
DM-850W, DM-850L,
DM-890, DM-940,
AM-890, AM-940 (DM
and AM Series
Illuminators)
Wavelength: 850 to 940 nm.
Eye illumination level: less than 1 mW/cm
²
at >450mm from illuminator.
Powered from camera via integrated cables.
Camera Ethernet
Cabling: Unshielded CAT5e or CAT6 cables up to 30 meters in length.
Requires host computer with supported Ethernet hardware capable of
gigabit speeds.
Power Supply
Specifications
GL camera: 12VDC, 2A external power supply with 2.5mm coaxial
(“barrel”) power connector (5.5
×
2.5
×
9.5mm).
OC camera head and FL illuminator: 3.6-5.6VDC@2A minimum. When
used with supplied power harness, DB-9 connector is required (contact SR
Research for pinout).
Power supply must have EN 60950, UL 950, CSA 22.2 No. 950, or other
equivalent safety approval, with LPS or Class 2 rating.
Operating conditions
10°C to 30°C, 10%-80% humidity (non-condensing)
For indoor use only.
Storage conditions
-10°C to 60°C, 10%-90% humidity (non-condensing). Allow to warm to
room temperature before unpacking or use after storage at temperatures
below 10°C to prevent condensation.
Safety Standards
IEC 60950-1:2005 (2nd Ed.) + Am 1:2009
IEC 60825-1:1993 + A1:1997 + A2:2001
IEC 62471:2006 (1st Ed.)
FDA 21 CFR laser products, under Laser Notice No. 50.
Fiber Optic Link (GL
and OC modules)
Class 1 laser product, IEC 60825, CFR 21
850 nm, multimodal duplex fiber, LC connector.
Electromagnetic
compatibility and
immunity
FCC Part 15, Subpart B: Class A unintentional radiators (see statement
below)
CISPR 11:1997 and EN55011:1998 -- Class A