- 5 -

Overheat preventive device activates at approx. 120

(no-liquid heating)

Safety functions

Safety unit

Operation

Causes

How to recover

Fuse

The fuse will blow to shut off 
power when overcurrent flows.

The power circuit was short-circuited 
or was subjected to overcurrent.

See P15 “7-2 How to replace 
fuses”.

Over heat 

preventive device 

(high limiter)

Shuts off power to stop control 
when the temperature in the 
bath reaches the set 
temperature for the overheat 
preventive unit. 

The unit is heating without any load. 
The temperature controller was 
malfunctioning and the temperature 
in the bath rose above the set 
temperature of the overheat 
preventive unit. 

See P13 “5-4 How to 
recover the overheat 
preventive unit”.

3-2  Alarm functions

The product has the following safety functions.
If you encounter with an abnormality, take appropriate

measures referring to P.14“Possible causes of troubles 
and solutions”.

Appears 
alternately with the 
liquid temperature

The measured temperature exceeds the upper limit setting of the 

temperature.

Check the upper limit temperature setting by referring to “P11  5.3  How to 

set the alarm function”. You clear the display with the SET key if the 
measured temperature is below the setting. 
If you cannot release the alarm, stop operation and contact your dealer or 
the nearest service center. 

Temperature upper limit 
alarm (control stop)

•Measured temperature is below the lower limit of the measurable 
temperature range (0

).

•Measured temperature exceeds the upper limit of the measurable 
temperature range (210

).

Remove any causes or heating or cooling in the unit. 
Then check the liquid temperature and the environmental temperature and 
take care to keep the temperature in the safety range. 
You can clear the display with the SET key if the measured temperature is 
within the safe temperature range 

0

210

). 

If you cannot release the alarm or the measured temperature is obviously far 
different from the actual temperature, immediately stop operation and contact 
your dealer or the nearest service center.

Outside the measurable 

temperature range alarm

(control stop)

Name / (operation)

Display

Causes and how to recover

Temperature sensor 
alarm (control stop)

The temperature sensor in the product short-circuited or had a 

disconnection.

When this alarm is displayed, immediately stop operating the unit and 
contact your dealer or the nearest service center. 

Power outage recovery 

alarm

Differs depending on 

the A-4 setting

Power was shut off during temperature control and power was turned on 

again.

You can clear the display with the SET key. 
You can set whether temperature control is stopped or continued during 
alarm display by setting ON/OFF for the power outage recovery function 
(See “P11  5.3 Setting the alarm function” ).
When you cannot clear the display with the SET key or the ON/OFF setting 
of the power outage recovery function is not apparently reflected, stop 
operation and contact your dealer or the nearest service center. 

Alarm functions

Appears 
alternately with the 
liquid temperature

Содержание SB-1300

Страница 1: ...signed to use the product safely while keeping its best performance Please keep this manual in a place easily accessible to every user IMPORTANT Be sure to read Safety precautions before use Tokyo Rikakikai Co Ltd 取扱説明書 Instruction Manual Combining with Rotary Evaporator ...

Страница 2: ...s recommended to follow the instruction to ensure the safe use of the product We have undertaken thorough verification concerning the possible occurrence of risk when using the product but prediction of all and every kind of risk is extremely difficult Namely cautions contained in this manual are not necessarily all of possible risks However if the product is operated according to the procedure de...

Страница 3: ...s are as shown below The warning labels should be read before using the product If warning labels become illegible due to damage on them replace them with new ones Contact us for spare labels Water Bath Model SB 1300 2 Warning display on the product CAUTION Hot surface Do not move the unit When liquid temperature is high ...

Страница 4: ...trol panel 3 2 Alarm functions 4 Installation 4 1 Installation environment 4 2 Connecting utilities Thank you very much for purchasing product Foreword Contents Package Contents This instruction manual describes the procedure of set up operations troubleshooting maintenance check up and disposal of SB 1300 which is water bath for rotary evaporator Be sure to carefully read this manual and understa...

Страница 5: ...ath with the cover heat remains in the bath and it gets hot Then the cover may be melted or ignited and it may result in burn or fire Specifically hot part Do not move the bath when the liquid temperature is high During use or for a certain time after use the edge of bath and liquid in the bath get hot If you move the bath in such case splashed liquid or the edge of bath may cause scald Edge of ba...

Страница 6: ...120H mm Container capacity Max 2ℓ Operating liquid temperature range 10 95 Operating environmental temperature range 2 5 35 External dimensions including handle W285 D295 H244 mm Mass 3 9kg Rated power supply power input AC230V 50Hz 6A Pollution degree 2 Over voltage category Ⅱ Operation at a terrestrial altitude Max2000m above sea level This product is a water bath specifically intended for heati...

Страница 7: ...l panel Power switch Bath cover Handle Overheat preventive device reset button Power cord Power supply inlet with Fuse 8A 2 Communication connector A special communication cable is used for connection with the vacuum control unit model NVC 3000 ...

Страница 8: ...d when an abnormal condition occurs RUN lamp Comes on when temperature control is started HEATING lamp Comes on while the heater is supplied with power RUN STOP key Used to switch start stop of temperature control UP key Each time this key is pressed the set temperature will rise by 1 The value increments continuously if it is kept pressed DOWN key Each time this key is pressed the set temperature...

Страница 9: ... of the measurable temperature range 0 Measured temperature exceeds the upper limit of the measurable temperature range 210 Remove any causes or heating or cooling in the unit Then check the liquid temperature and the environmental temperature and take care to keep the temperature in the safety range You can clear the display with the SET key if the measured temperature is within the safe temperat...

Страница 10: ...talling this product Place that the driving assembly or glass parts will not come into contact if the evaporator jack is lowered Place free of flammable gas liquid or solid materials in the vicinity of the product Place where the ambient temperature can be kept within a range of 5 35 Place free from condensation Place with less humidity and free from splashing water Place with minimum dust Place f...

Страница 11: ... Check the product type as well as the voltage phase and capacity of power supply to be connected Power supply to be connected to the product is as shown in the right 2 Check the receptacle of installation place Prepare an outlet with a ground terminal Specification of Power Cord 4 2 Connecting utilities Warning Confirm the voltage phase capacity and the type of receptacle of power supply Wrong co...

Страница 12: ...l or the flask may break 4 Gently pour water or hot water in the specimen flask to the correct level to the capacity of the flask with the flask put in the bath If water amount is excessive water may flow over the specimen flask while it is rotating Keep the water level at least 2cm below the top surface of the bath 5 Raise the evaporator jack to the highest point Also refer to instruction manual ...

Страница 13: ...correct amount of water is in the bath 1 Turn the power switch ON The whole display comes on indicates the model Sb the mode switches to the measurement mode and displays the present liquid temperature Temperature indication on 2 Pressing the SET key switches to the setting input mode Temperature indication flashes The temperature is set to 25 at the time of factory shipping The setting in the las...

Страница 14: ...ng do not place it upside down or on its side 7 Rotate the specimen flask to start condensation Raise the rotation speed only slowly to avoid the water from spilling over or ruffling Take care not to burn yourself since hot water may ruffle while the flask is rotating Do not put the cover on the bath The cover may burn from the heater heat And overheating may cause inside the bath hot and a malfun...

Страница 15: ...ture limit setting range 10 120 The temperature is set at the upper limit temperature 120 at the time of factory shipping 5 Pressing the SET key switches to the alarm function setting mode and the setting of the upper temperature limit is completed A 0 indication on 6 Press the SET key long for about 5 seconds to return to the measurement mode 1 Press the SET key longer in the measurement current ...

Страница 16: ...recovery from power outage The setting is on at the factory shipping 2 Use the UP key and the DOWN key to set ON OFF for the power outage recovery function 3 Pressing the SET key switches to the alarm function setting mode and the setting of the power outage recovery function is completed A 4 indication on 4 Press the SET key long for about 5 seconds to return to the measurement mode When the powe...

Страница 17: ...ure that the inside of the bath has sufficiently cooled down and then press the reset button of the overheat preventive unit 3 Turn the power switch ON 4 The display comes on again and temperature control starts RUN lamp on If the temperature in the bath when the power is turned on by turning the power switch ON after resetting the overheat preventive unit the heater will be activated automaticall...

Страница 18: ... RUN STOP key is not pressed Press the RUN STOP key to start temperature control Temperature rises above the set temperature HEATING lamp is off Set temperature is low Start condensation and make sure that the temperature falls The environmental temperature is higher than the measured temperature Use the product at a set temperature higher than the environmental temperature 5 35 SSR is malfunction...

Страница 19: ...one may cause overheat of the bath bottom and damage the product Caution Do not scratch the bath Inner surface of the bath is Teflon coated for being cleaned efficiently Scratching the inner surface may damage the product Never attempt to disassemble the product The unit contains parts with high voltage applied or may become hot and disassembly may cause an electrical shock or an injury 7 2 Replac...

Страница 20: ...ntegrated bath Stainless steel aluminum ceramics Bath cover PET containing glass fibers Body case PBT containing glass fibers Handle PP Metal sheet Aluminum iron stainless steel Disposal of Products Disposal of product or part must be done according to the specified disposal method Principal components parts and disposal method We ask you to discard packing materials after classifying them by mate...

Страница 21: ...k first by referring to the page on troubleshooting to see if this is actually a trouble 2 If the product remains unsatisfactory even after checking contact the shop from which the user has purchased the product or the service center described in the manual and request repair 4 After expiration of the guarantee period the charged repair will be made at the customer s request ...

Страница 22: ...dicated for connecting with N 1200 It cannot be used for connecting with other evaporators Time delay Fuse 1 241088 8A 2 2 245373 10A 1 230V Power Cord B Type 1 2 3 4 3 245372 10A 1 230V Power Cord C Type 4 245371 10A 1 230V Power Cord O Type 5 242518 10A 1 5 5 No Code No Std Qty No Code No Std Qty No Code No Std Qty No Code No Std Qty Connection using an N SB connection cord Service outlet N 1200...

Отзывы: