- 8 -
5
Do not put the liquid other
than water into the bath.
This product is intended for water (fresh water) only.
If liquid other than water is used, it may cause fire
or failure.
(1) Raise the evaporator jack to the highest point.
(2) Place the bath so that the specimen flask of the
evaporator is at the center of the bath.
(3) Slowly lower the evaporator jack to determine a point
where the flask, the rotary joint or the joint clip does
not come into contact with the bath. (Turn the flask by
hand to make sure that it does not come into contact
with any part.)
※
Do not lower the evaporator jack while the bath
handle is at the raised position.
Otherwise, the evaporation may fall or the flask
may break.
(4) Gently pour water (or hot water) in the specimen flask
to the correct level to the capacity of the flask with the
flask put in the bath.
※
If water amount is excessive, water may flow over
the specimen flask while it is rotating. Keep the
water level at least 2cm below the top surface of
the bath.
(5) Raise the evaporator jack to the highest point.
※
Also refer to instruction manual for rotary
evaporator (N-1300,N-1200 series).
FREE
LOCK
Operation
5-1 Preparation for operation
Warning
1. Installation of the bath
2.
Connecting the power cord
(1)Insert the power socket into the power inlet at the back
of the bath.
(2)Make sure that the power switch at the side of the bath
is OFF and then insert the power plug into the outlet.
OFF state
ON state
Power switch
Specimen flask
Jack
Bath
Handle
Power socket
Power inlet
Power plug
To the outlet