2
- 2 -
2-1 Application
2-3 Specifications
For water only.
This product is intended for water only.If you use
liquid other than water (pure water), it may cause fire
or failure.
Caution
Product Name
Constant temperature water bath (Water bath
)
Model
SB-1300
P
er
fo
rm
an
ce
Temperature set range
Room temp.+ 5
℃~
90
℃
Temperature Adjustment Accuracy
※
1
±
1
℃(
during stirring
)
Temperature Set Accuracy
※
1
±
5
℃(
during stirring
)
Temperature Indication Range
0
~
210
℃
F
u
n
ct
io
n
s
Temperature Control
P.I.D control
Temperature Indication
・
Setting
Digital display
(
Minimum digit1
℃)・
sheet key input
Safety Functions
・
Overheat preventive device (fixed temperature
・
liquid bloating manual recovery type
)
・
Fuse
・
Heat insulation protective bath cover
・
Self diagnostics function
Sensor disconnection/short-circuit alarm
Outside the measurable temperature range alarm, upper limit alarm
Power outage detection, watch dog
C
o
n
fig
u
-
ra
tio
n
Heater
1400W
For water only
(
For heating can body of the bath
)
Temperature sensor
Pt sensor
Bath cover
PET
(
with glass fiber
)
S
ta
n
d
ar
d
Material of bath
SUS304
Bath capacity
Approx. 4.3L
Effective size in the bath
φ
220
×
120H
(
mm
)
Container capacity
Max 2
ℓ
Operating liquid temperature range
10
~
95
℃
Operating environmental temperature range
※
2
5
~
35
℃
External dimensions (including handle)
W285
×
D295
×
H244
(
mm
)
Mass
3.9kg
Rated power supply
・
power input
AC230V
50Hz
・
6A
Pollution degree
2
Over voltage category
Ⅱ
Operation at a terrestrial altitude
Max2000m above sea level
This product is a water bath specifically intended for
heating specimen flask of Rotary Evaporator to conduct
condensation efficiently.
For details of setting methods, see Page 9 “5-2 How
to Operate”.
Product Outline
Warning
Never attempt to modify the product.
Operate the unit for the specified purpose only.
An electric shock or a malfunction may result if the
product is modified or used for any purposes other
than that specified.
2-2 How to use
The temperature of the bath can be set depending on purposes.
With using control panel, the temperature can be set.
Preset temperature can be changed and operation can be
stopped if needed as long as the operation is underway.
※
1 Performances have been measured room temperature of 20
℃
; rated power voltage, 50Hz, no-load, during
stirring (1L flask)
※
2 Use the unit at an environment of room temperature
:
20
℃
±
5
℃
; humidity
:
60
%
or lower when a cool
water
(
5
℃
or higher
)
.
Allow the inside of the unit to completely dry after use.