SCPI Command Reference
Tunable Laser Source
93
T200S Specific Commands
OUTPut[:STATe]?
Description
This query returns the state of the laser output.
At *RST, this value is set to ON.
Syntax
:OUTPut[:STATe]?
Parameter(s)
None.
Response Syntax
<state>
Response(s)
state:
State of the laser output:
³
0 : the laser output is disabled, which means that the source
emission is disabled.
³
1 (default): the laser output is enabled.
Example(s)
OUTP ON
OUTP? returns 1 (open)
See Also
OUTPut:MONitor1|2[:STATe]
Description
This command controls the activation of the output signal
generation on the specified monitoring BNC connector located on
the rear panel:
³
MONitor1 is the
λ
out
BNC connector (for spectrum
monitoring).
³
MONitor2 is the
Power out
BNC connector (for power
monitoring).
During the execution of this command, bit #1 (SETTling) is set in
the Operation Status Condition register (see
At *RST, this value is set to OFF.
Syntax
:OUTPut:MONitor1|2[:STATe]<wsp><state>
Parameter(s)
state
:
State of the specified monitoring output. The allowed values are:
³
OFF|0 (default): the output is closed, which means that no
signal is output from the BNC port.
³
ON|1: the output is open.
Example(s)
OUTP:MON:STAT OFF
OUTP:MON:STAT? returns 0 (closed)
See Also