Remotely Controlling the T200S
48
T200S
Communication Principle
Operational / Questionable Status Reporting
Operational Status Condition Register
The following table shows the content of the Operational Status Condition Register:
Bit
Weight
Meaning
15
32768 Always zero.
14
16384 PROGram running: a user-defined programming is currently in the run
state.
13
8192
INSTrument Summary Bit: One of n multiple logical instruments is
reporting OPERational status. (NOT USED)
12
4096
Shutting Down: The instrument is performing a shutdown.
11
2048
Warming Up: the instrument is warming up.
10
1024
Always zero.
9
512
INITializing: the instrument is currently performing initialization.
8
256
REFerencing: the instrument is currently performing a reference.
7
128
CORRecting: the instrument is currently performing a correction.
6
64
Waiting for ARM: the instrument is in a “wait for arm” state of the
trigger model.
5
32
Waiting for TRIG: the instrument is in a “wait for trigger” state of the
trigger model.
4
16
MEASuring: the instrument is actively measuring.
3
8
SWEeping: a sweep is in progress.
2
4
RANGing: the instrument is currently changing its range.
1
2
SETTling: the instrument is waiting for signals it controls to stabilize
enough to begin measurements.
0 (LSB)
1
CALibrating: the instrument is currently performing a calibration.