SCPI Command Reference
116
T200S
T200S Specific Commands
[SOURce]:WAVelength[:CW|FIXed]?
Description
This query returns a value indicating either the current absolute
wavelength or the minimum/maximum/default absolute
wavelength of a non-swept signal.
At *RST, this value is set to DEFault.
Syntax
[:SOURce]:WAVelength[:CW|FIXed]?<wsp>
[MAXimum|MINimum|DEFault]
Parameter(s)
³
MAXimum
The query returns the maximum possible wavelength value
for the laser model.
³
MINimum
The query returns the minimum possible wavelength value for
the laser model.
³
DEFault
The query returns the default wavelength value for the laser
model.
Response Syntax
<wavelength>
Response(s)
wavelength:
Value in meter corresponding to the current wavelength or the
minimum/maximum/default wavelength, depending on the
specified query parameters.
Example(s)
SOUR:WAV:FIX 1570 NM
SOUR:WAV:FIX? returns 1.570000E-006
See Also
[SOURce]:FREQuency[:CW|FIXed]?
[SOURce]:OPTimization:OLINewidth[:STATe]?
[SOURce]:WAVelength[:CW|FIXed]