SCPI Command Reference
86
T200S
T200S Specific Commands
CALibration:DATA:TIME?
Description
This query returns the time of the last successful wavelength
referencing operation.
Syntax
:CALibration:DATA:TIME?
Parameter(s)
None.
Response Syntax
<hour>,<minute>,<second>
Response(s)
hour:
Integer corresponding to the hour of the current calibration in 24
hour format, in the range 0 to 23.
minute:
Integer corresponding to the number of minutes in the range 0 to
59.
second:
Integer corresponding to the number of seconds, in the range 0 to
59.
Example(s)
CAL:DATA:TIME? returns 14,57,09
See Also
CALibration:WARMup:ABORt
Description
This command stop the warmup of the instrument.
This command is an event and has no associated *RST condition
or query form.
Applicability
This command is only valid if bit #11 (WarmingUp) of the
Operation Status Condition register is set (see
Syntax
:CALibration:WARMup:ABORt
Parameter(s)
None
Example(s)
CAL:WARM:ABOR
See Also