SCPI Command Reference
Tunable Laser Source
125
T200S Specific Commands
[SOURce]:WAVelength:SWEep:STOP?
Description
This query indicates either the current or the
minimum/maximum/default stop wavelength of a swept signal.
At *RST, this value is set to MAXimum.
Syntax
[:SOURce]:WAVelength:SWEep:STOP?<wsp>
[MAXimum|MINimum|DEFault]
Parameter(s)
³
MAXimum
The query returns the maximum possible stop wavelength for
the laser model.
³
MINimum
The query returns the minimum possible stop wavelength for
the laser model.
³
DEFault
The query returns the default stop wavelength for the laser
model.
Response Syntax
<stop>
Response(s)
stop:
Floating point number in meter corresponding to the current stop
wavelength or the minimum/maximum/default stop wavelength,
depending on the specified query parameters.
Example(s)
SOUR:WAV:SWE:STOP 1490 NM
SOUR:WAV:SWE:STOP? returns 1.490000E-006
See Also
[SOURce]:FREQuency:SWEep:STOP?