MFC
Functions
4.
59
7
Rinse off the escaped electrolyte with water.
Please note
The membrane cap (E) must be completely screwed (hand-tight)
onto the electrode shaft (B), so that no gap remains between the
two! After a run-in period of about one hour, the membrane sensor
is sufficiently run-in for an initial calibration to take place.
Calibration should be repeated after one day.
Inserting into the flow-through
adapter
1
Insert the membrane sensor through the cover into the flow-
through adapter until it is resting mechanically on the inflow
mating connector and therefore receives a good flow. It may
be necessary to turn the flow-through adapter cover to set it in
the right position relative to the inflow mating connector.
Please note
Remove air bubbles from the membrane by lifting the membrane
sensor. They interfere with the measurement!
2
Connect the measuring signal cable to the measuring device.
Please note
If the measuring sensor is installed this way, it will function reliably
for approx. three to six months.
Storing the membrane sensor
1
Lift elastomer seal (I) and only then screw off the membrane
cap (E).
2
Remove G-holder (K) (with tweezers) and rinse together with
the membrane cap (E) and electrode finger (D) with clean
(distilled) water.
3
Carefully dry the electrode finger (D) with absorbent paper.
4
Leave the membrane cap (E) to dry in a dust-free place.
5
Screw the dry membrane cap (E) loosely onto the electrode
shaft (B).
6
Place G-holder (K) loosely into the membrane cap, do not fit
onto the electrode finger (D).
Please note
The membrane (G) may not touch the gold working electrode (L).
Restarting
See "Commissioning“.
Содержание WALLACE & TIERNAN MFC
Страница 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Страница 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Страница 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Страница 225: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 225 DEPOLOX 5 drawing non pressurised version W3T170031 W3T170031...
Страница 229: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 229 DEPOLOX 5 drawing pressurised version W3T170032 W3T170032...
Страница 233: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 233 VariaSens drawing non pressurised version W3T158742 W3T158742...
Страница 237: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 237 VariaSens drawing pressurised version W3T158756 W3T158756...
Страница 247: ...MFC Certificates 10 247 10 Certificates...
Страница 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 249: ...MFC Certificates 10 249...
Страница 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...
Страница 266: ...266 WT 050 580 000 DE IM 0714 Index MFC 12...