MFC
Operation
5.
157
5.4.4
pH calibration
Please note
During pH calibration the buffer solution and the sample water
should have the same temperature. If there is a difference in
temperature of > 5 °C, first enter the temperature of the buffer
solution in the "Calibration" - “pH (1) ... pH (4)“ menu for the
measurement you wish to perform (Display 2.1.2 “Calibration -
Menu 2.1" on page 147) under "Cal. at temp.".
For buffer solutions, refer to “Automatic calibration" on page 200.
pH-7 alignment
1
Starting from the basic display in the main menu select the
"Calibration" menu.
2
Select the menu item “pH (1) ... pH (4)“ for the measurement
you wish to perform.
The "Calibr. pH" window (Display 2.1.2) appears on the
graphic display.
3
Select the "Cal. pH7" parameter.
4
Place one of the supplied beakers into the bottom clip and fill
with the buffer solution "pH 7.00" or a bag with buffer solution
"pH 7.00" into the bottom clip.
5
Pull or unscrew the pH sensor from the lid of the cell body.
6
Dip the pH sensor through the top clip at least 2 cm deep into
the buffer solution and move slightly until the indicated pH
value remains constant.
7
Enter the pH value associated with the buffer temperature
using the arrow keys or or leave pH 7.00.
8
Save the value by pressing "Enter".
Slope alignment
9
Remove the buffer solution "pH 7.00" from the bottom clip.
10
Wash the sensor in distilled water to prevent carry-over of the
buffer solution.
11
Select the parameter "Cal. pH" and confirm the selection.
12
Place a second beaker into the bottom clip and fill with the
buffer solution "pH 4.65" or clamp a bag with buffer solution
"pH 4.65" into the bottom clip.
Please note
If buffer solutions other than those stated are used, the pH value
of the buffer solution must be lower than pH 6 or higher than pH 8 .
Содержание WALLACE & TIERNAN MFC
Страница 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Страница 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Страница 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Страница 225: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 225 DEPOLOX 5 drawing non pressurised version W3T170031 W3T170031...
Страница 229: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 229 DEPOLOX 5 drawing pressurised version W3T170032 W3T170032...
Страница 233: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 233 VariaSens drawing non pressurised version W3T158742 W3T158742...
Страница 237: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 237 VariaSens drawing pressurised version W3T158756 W3T158756...
Страница 247: ...MFC Certificates 10 247 10 Certificates...
Страница 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 249: ...MFC Certificates 10 249...
Страница 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...
Страница 266: ...266 WT 050 580 000 DE IM 0714 Index MFC 12...