DEPOLOX
®
5 E flow cell
Installation
4.
53
4.4.14
Fitting the sens (optional)
Please note
The membrane sensors, combined measuring and reference
electrodes and electrodes must be prepared accordingly. Please
follow the appropriate instructions for the sensors! On the pres-
surised version it is not possible to install a membrane sensor.
Depending on requirements, insert or screw in the sensors into the
mount hole on the cover of the cell body.
Please note
With the pressurized version, the sensors must be screwed in or
secured to prevent them from being pushed out.
Proceed as follows:
1
On the non-pressurized version, remove the cap from the
mounting hole on the cell body cover.
2
On the pressurized version, unscrew the protection cap.
3
Take the pH and redox combined measuring and reference
electrode out of the KCl tank with stand.
Please note
Keep the KCl tank with stand for later use or storage.
4
With the conductivity sensor, remove the blind plug and O-
ring.
Содержание WALLACE & TIERNAN DEPOLOX 5 E
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0817 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Страница 45: ...DEPOLOX 5 E flow cell Installation 4 45 Example for sample water take off Only operate with sample water inlet...
Страница 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Страница 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Страница 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...