DEPOLOX
®
5 E flow cell
Installation
4.
41
4.4.5
Removing or fitting the housing cover
1
Remove the housing cover of the DEPOLOX
®
5 E flow cell. To
do this, press both unlocking buttons on the top of the housing
and carefully remove the housing cover forwards.
2
Refit and engage the housing cover of the DEPOLOX
®
5 E
flow cell. To do this, place the housing cover against at the bot-
tom of the housing and carefully push it upwards until the
housing cover engages with the unlock buttons.
Figure 1
Top view of housing cover
A
Unlocking buttons
4.4.6
Connecting the sample water inlet
Please note
Do not install water carrying lines made of copper. These would
distort the measurements.
When connecting the sample water inlet, observe the following:
•
Sample water inlets are classified according to whether they
have a hose connection or are fitted with rigid piping.
•
The pressure in the sample water inlet must always be within a
range of 0.25 to 3.0 bar. The pressure in the sample water
inlet must generally be 0.25 bar higher than in the sample
water outlet.
•
At an admission pressure below 0.25 bar, an optional pressure
booster pump must be used. See chapter 4.4.7 – “Example for
sample water take-off with a booster pump”.
•
If the admission pressure exceeds 3.0 bar, an optional pres-
sure reducing valve must be used. See chapter 4.4.7 – “Exam-
ple for sample water take-off”.
•
To prevent long loop lag times, ensure that the lines in the
sample water inlet are as short as possible.
•
An external strainer with a mesh width of 0.5 mm is provided
for the sample water inlet.
Figure 2
Cross-section of DEPOLOX
®
5 E flow cell, non-
pressurized
A
Sample water inlet with shut-off ball valve
B
Sample taking unit (drain on the drain screw)
C Sample water outlet
A
A
A B C
Содержание WALLACE & TIERNAN DEPOLOX 5 E
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0817 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Страница 45: ...DEPOLOX 5 E flow cell Installation 4 45 Example for sample water take off Only operate with sample water inlet...
Страница 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Страница 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Страница 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...