52
WT.050.810.040.DE.IM.0817
Installation
DEPOLOX
®
5 E flow cell
4.
4.4.13
Connecting sensors with DFMe electronics module
Proceed as follows:
1
Insert the multi-sensor (A) in the flow control valve (B).
2
Connect sensor cable DFMe-DES (E) to the 3-electrode mea-
suring cell (C). To do this, screw cap (D) on the
3-electrode measuring cell (C) counterclockwise up to the
mark (see Figure 10, item K) and take it off. Feed the sensor
cable through the cap and tighten with the M12 cable union
(L). Connect the signal cable, observing the correct colors
(see Figure 11 and lower table). Screw in the cap clockwise up
to the mark.
3
Insert or screw the LED glow stick (F) into the mounting hole in
the cell body cover.
4
Connect the pH- and redox sensor cable plug (G, H) with the
pH and redox combined measuring and reference electrode.
5
Plug the CAN connection cable into the CAN connection
socket (J) and with connect with 700 P electronics module or
700 M electronics module (CAN socket).
Figure 12 Cross-section of DEPOLOX
®
5 E flow cell, non-pres-
surized version
Figure 13 Cross-section, DFMe electronics module
A
E
F G
H
I
J
A
Multi-sensor
B
Flow control valve
C 3-electrode measuring cell
D Cap
E
Sensor cable DFMe-DES
F
LED glow stick
G pH sensor cable (optional)
H Redox sensor cable (mV) (optional)
I
CAN extension socket (optional)
J
CAN connection socket to
electronics module
A
B
C
D
Figure 10 Detail: cap
Figure 11 Section of 3-elec-
trode measuring cell
WRK
Working electrode
(Figure 11, item M)
Red point
Red cable
CNT
Counter electrode
(Figure 11, item N)
Blue point
Blue cable
Ref
Reference electrode
(Figure 11, item O)
White cable
K
L
M
N
O
Содержание WALLACE & TIERNAN DEPOLOX 5 E
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0817 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Страница 45: ...DEPOLOX 5 E flow cell Installation 4 45 Example for sample water take off Only operate with sample water inlet...
Страница 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Страница 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Страница 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...