56
WT.050.810.040.DE.IM.0817
Installation
DEPOLOX
®
5 E flow cell
4.
4.6.2
Taking sensors out of operation
Please note
Please follow the appropriate instructions for the sensors!
For the pH or redox combined measuring and reference electrode,
proceed as follows:
1
Pull out or unscrew the sensors.
2
Fit the pH or redox combined measuring and reference elec-
trode into the KCl tank and stand with KCI solution.
3
Store the sensors in a frost-free place.
Figure 17 pH and redox combined measuring and reference
electrode with stand
A
Sensor
B
Sealing cap
C O-ring
D Tank
E
Stand
4.7
Recommissioning
1
Prepare the sensors. See appropriate sensor instruction
manual.
2
Fitting the sensors and connect the sensors with DFMe elec-
tronics module. See chapter 4.4.14 “Fitting the sens (optional)”
and chapter 4.4.13 “Connecting sensors with DFMe electron-
ics module”.
3
Filling electrode-cleaning sand, see chapter 4.4.12 “Filling
electrode cleaning sand”.
4
For recommissioning, see instruction manual “700 P electron-
ics module” or “700 M electronics module”.
B
C
E
A
D
Содержание WALLACE & TIERNAN DEPOLOX 5 E
Страница 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 810 040 DE IM 0817 DEPOLOX 5 E flow cell Please note Original instruction manual Simone...
Страница 8: ...8 WT 050 810 040 DE IM 0817 Introduction DEPOLOX 5 E flow cell 1...
Страница 45: ...DEPOLOX 5 E flow cell Installation 4 45 Example for sample water take off Only operate with sample water inlet...
Страница 82: ...82 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 83: ...DEPOLOX 5 E flow cell Declarations and certificates 7 83 7 2 Certificate of CSA...
Страница 84: ...84 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 85: ...DEPOLOX 5 E flow cell Declarations and certificates 7 85...
Страница 86: ...86 WT 050 810 040 DE IM 0817 Declarations and certificates DEPOLOX 5 E flow cell 7...
Страница 90: ...90 WT 050 810 040 DE IM 0817 Index DEPOLOX 5 E flow cell 8...