Instruction Manual MM850404 ed.05
49
2 - Connect the infrared interface adapter to the serial port of the PC with the supplied cable (note back the serial port
used: COM1 , COM2 or other).
3 - Connect the interface to the instrument.
4 - Run the ST167Flash program you have installed.
5 - Select the same serial port that you noted back.
6 - Select the Baud rate to 115200 Baud (if the PC owns an UART 16550 type), otherwise select a lower Baud rate.
7 - Press the three dotted button in order to load the firmware file (.HFW extension) normally located on a 3 1/2" diskette.
A version number of the firmware being discarded is shown in the dedicated field.
8 - From instrument OFF press and continue to press the F1 key while pressing the ON/OFF key in order to set the
instrument in firmware upgrade status.
9 - Click on the 'Program' button to start the upgrading phase and wait until a termination message appears.
NOTE:
I
F DATA TRANSFER FAILS
,
PLEASE TURN OFF THE INSTRUMENT AND REPEAT THE PROCEDURE FROM STEP
8.
9.3
Smoke Pump Maintenance
Gas-tight checking
To have a correct measurement the pump should be gas-tight. Turn the screw and close the probe with your finger. Pull
the plunger for 3 cm and verify that the gasket does not leak from the vacuum. If a leak is found the gasket needs to be
lubricated. When you push the plunger the valve must close immediately to have a 1.9/2.0 litres of air with 10 shoots.
The correct smoke index can be obtained only if the pump works correctly.
Service
1. Clean and lubricate the cylinder the stem and the gasket every 30 measurements.
2. Wash using alcohol if the valve is not gas-tight
3. Clean the probe and the paper support
4. Check the pump for gas-tight
5. Use only clean paper filters and a readable scale index table
9.4
Error messages
The instrument
performs the autodiagnostic procedure when turned on. When measuring it checks also for anomalous
conditions. If analyzer founds trouble, it displays an error for the related parameters. Printing the report, the error
messages will be printed. Sometimes these messages correspond to a unrecoverable problem and it is necessary to
return the instrument to Eurotron for maintenance. Frequently these messages are referred to a recoverable trouble and
can be solved by yourself.
The following table shows the possible instrument error messages:
Error message
Problem
Solutions
Calc Err
Parameter calculation error.
The instrument cannot perform the
calculations. Check and solve the others
error conditions.
Probe Er
The external Rtd probe is
disconnected or damaged.
Connect or replace the external temperature
probe.
RJ Err
Internal temperature reference
damaged.
Return the instrument to Eurotron for
maintenance
Over
Overrange condition
The problem is solved automatically when
measure return inside the range
Under
Underrange condition
The problem is solved automatically when
measure return inside the range
--------
No calculations available
The instrument cannot perform the
calculation. Check and solve the others