Instruction Manual MM850404 ed.05
24
Press
[
ESC
] to return to the measure page.
6.1.4
Pressure/Draft Measurements
The 2000 series
has an internal differential pressure sensor installed. With this sensor, you can execute the
pressure/draft measurements by using or the standard probe or the special dual probe.
Standard single hose probe (TG models):
The standard probe has one gas collector for gas sampling. Usually you have the hose terminal connected to the
instrument gas inlet; when you have to do the draft measurements you have to move the connector to pressure inlet
(INLET) connector.
Special dual hose probe (TGD models):
This special probe model has two gas collectors: one for the flue gas analysis and one for the draft measurement. When
you connect the probe to the analyzer you have to insert both gas inlet (INLET) and pressure inlet (P1) connectors.
Connect the hose with line filter to gas inlet connector.
CAUTION
I
F A SINGLE TUBE GAS SAMPLING PROBE
(TG
MODEL
)
IS USED FOR
P
RESSURE
/
DRAFT MEASUREMENTS
,
BE SURE THAT THE PROBE
AND HOSE ARE CLEAN AND DRY BEFORE CONNECTING TO THE PRESSURE INPUT
.
Press the [
MENU
] key until the following page is displayed:
Press the [
ENTER
] key to show the following page. The internal pump will stop.
NOTA:
DO
NOT
INSERT
THE
PROBE
IN
THE
CHIMNEY.