Instruction Manual MM850404 ed.05
28
To print the data analysis follow the procedure:
Wait for the stabilization of the measure on the display. Press the [
]
key.
The display shows the last setting for number of copies to be printed and
the document format.
Check that the internal printer is enabled. Press the [
MENU
] key and verify
that it dispays the “
PRINTINT
” message.
If you do not need any changes press the [
] key again. The
instrument will print the report as programmed
If you need any changes, read the procedure on paragraph 6.3.3.
Press
[
ESC
] to return to the measuring page.
6.3.2 External
Printer
The 2000 series
has a standard IR port compatible with your HP 82240B wireless thermal printer. The printer is external
to the instrument and the power supply si separated from the analyser. No connection cable are necessary: you can
nearing the printer IR led to the IR led of the analyser.
To print the analysis data follow the procedure:
Wait for the measurement to stabilize on the display. Press the [
] key.
The display shows the last setting for number of copies to be printed and the document format.
Check that the internal printer is enabled. Press the [
MENU
] key and verify that it dispay the “
PRINTEXT
” message.
If you do not need any changes press the [
] key again. The instrument will print the report as programmed
If you need any changes, read the procedure on paragraph 6.3.3.
Press
[
ESC
] to return to the measure page.
6.3.3 Printer
Setting
From the printer menu you can change the printer setting: the format of the report (short or complete form), the printer
enabled (internal or external) and the number of the copies. From this menu you can also perform the form feed also.
Press the [
] key to enter in the printer menu.
the display shows the last setting for number of copies to be printed and the document format.
Press the [
MENU
] key until the parameter to be changed is displayed.
Press the [
X
] key to change the value of the displayed parameter. Press the [
ENTER
] key to store the changes.
Press the [
MENU
] key to change the parameter.
Press the [
] key to print the report.
Press
[
ESC
] to return to the measure page.
6.3.4 Header
Setting
Using the instrument keyboard, you can set the report header (4 text rows of 16 characters each) with your data.
Adopt the following procedures:
Press the [
MENU
] key from measuring page until the following page is displayed:
Press the [
ENTER
] key. Press the [
MENU
] key until the following page is displayed: