Page 28
Description and elimination of faults
Fault displayed
Possible cause
Remedy
Dispenser does not
deliver
1. Bottle is empty
2. Pump tubes kinked (e.g.
underneath the pump
head)
3. Pump intake tube stuck to
side of bottle (when bottle
is nearly empty)
4. Wrong (high viscosity)
liquid
1. Insert new full bottle
2. Check tubes and eliminate
kink, if necessary
3. Check tube and position
intake tube in middle of
bottle, change bottle, if
necessary
4. Check liquid for
suitability before use
Dispenser delivers too
little
1. High viscosity liquid
used
2. Dispenser is set to small
dose e.g. level 1 =
approx. 1.6 ml
1. Change liquid
2. Set to a higher dose level s.
chapter 6
Pump runs stiffly
1.
Pump is broken, to old
2. High viscosity liquid is
in use
1. Change pump (preferably
every 3 to 4 months)
2. Check medium for
suitability before use
Sensor does not work
even if a hand is held
under it
1. Trigger distance is too big
2. Trigger distance has
become maladjusted
3. Sensor electronics defect
1. Bring hand closer to
dispensing tip
2. Re-adjust distance see
chapter 10
3. Contact dealer to discuss
further steps to be taken
Bottle
“
wobbles
”
in the
dispenser or falls off
when opening the
dispenser
Bottle fasteners not fixed;
Push bottle fastener
upwards (towards the base
of the bottle) until the bottle
sits firmly,
For 1,000 ml dispenser only:
center the additional variable
fastener precisely over the
neck of the bottle
Dispenser cannot be
closed properly
1.
Key is still inserted in lock
2.
Tube is clamped
1.
Lift/remove key, close
front cover properly
2.
Remove tube, close front
cover properly