21. CHANGING THE FUSES
WARNING: Do not perform the following operation in the presence of the patient.
Step 1
Open the fuse box cover
using a screwdriver
Step 2
Take out the fuse box
(use a screwdriver to
release it)
Step 3
Remove the blown fuse
from its seat and replace
it with an identical one,
as shown in the table
below and on the
instrument label.
Step 4
Push the fuse box
carefully back into
position
Fuse type
Fuse value
20 x 5 mm
T 2.5 A L 250 V anti-surge