8. Service and recalibration
E
08
Blood pressure gauges must be calibrated and marked CE when they are put on the
market.
We r
ecommend f
or
t
he doct
or
’
s and t
he pat
i
ent
’
s secur
i
t
y t
hat
t
he i
n
strument be
checked and recalibrated at least every 2 years either by the manufacturer or by an
authorised workshop.
Instruments which deviate from zero position in the range of 0 - 4 mmHg should be
subject to immediate recalibration. Instruments without seal (for example after a
repair) are considered as uncalibrated and have to be recalibrated.
The date of manufacture is shown by the first two numbers of the serial number.
A recalibration of the model SWITCH is not intended. The faulty manometer housing
can be replaced anywhere (even by unskilled people) by a newly calibrated one
especially provided for this purpose by the producer (see below).
We emphasize that only original ERKA. parts should be used for any spare or
repair. A combination or attachment of ERKA. products with other makes is not
possible.
ERKA. Switch:
easy adjusting from left to right hand use
Quick exchange of manometer housing.
Labels in 8 different colours.
In addition labels can be marked individually
.
We trust that working with your ERKA. blood pressure gauge
will always meet with your satisfaction.