9-5
9.2
Visual and acoustic signals
STATUS
visual
acoustic
HF off
Power switch on
•
Operating error on the front panel
••
••
••
HF generator on
••
••
LF leakage current greater than 0.05 mA
••
Error No.5
••
••
HF leakage current greater than 150 mA, but less than 300 mA
••
red
HF leakage current greater than 300 mA
••
red
Error No. 17
••
Output error
••
red
••
••
Max. activation time limit reached
••
red
Max. activation time limit exceeded too long
••
red
••
••
No neutral electrode connected to the unit
••
red
Error No.13
••
••
Single-surface electrode connected to the unit
•
green
Dual-surface neutral electrode connected to unit and contact
resistance between the contact surfaces less than 120 ohms
•
green
Dual-surface neutral electrode not applied to patient or contact
resistance between the contact surfaces and patient greater than 120
ohms
••
red
Error No.13
••
••
HF current relative to the contact resistance between the contact
surfaces of a dual-surface neutral electrode and patient too great
••
red
Error No.19
••
4
times
HF partial currents of the contact surfaces of a dual-surface neutral
electrode different
••
red
HF partial currents of the contact surfaces of a single-surface neutral
electrode too different
••
red
Error No.21
••
••
Interruption of the line between unit and neutral electrode or between
patient and dual-surface neutral electrode during HF activation
••
red
Error No.13
••
••
Auto Stop for Auto Coagulation
••
••
Activation error
••
••
••
Error message upon error recognition during self check
•
•
•
•
These signals are produced regardless of whether the unit is activated or not activated.
••
These signals are produced only if the unit is activated.
Содержание ERBOTOM ICC 350 M
Страница 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Страница 2: ......
Страница 8: ...1 2...
Страница 10: ...2 2...
Страница 18: ...3 8...
Страница 19: ...7 8 9 10 International Standard...
Страница 20: ......
Страница 42: ...4 22...
Страница 69: ...9 11 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 70: ...9 12 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 71: ...9 13 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 72: ...9 14 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 77: ...9 19 AUTO CUT MODE Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 79: ...9 21 SOFT COAGULATION AUTO BIPOLAR Power output as a function of the power limitation RL 125 ohms...
Страница 80: ...9 22 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms Forced 1 2 3 4...
Страница 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Страница 88: ...11 2...
Страница 94: ...12 6...
Страница 96: ...13 2...
Страница 98: ...14 2...
Страница 100: ...15 2...