The ERBOTOM ICC 350 is equipped with 11 memory locations for
storage of various basic settings. In this way, the settings required for
every application of high-frequency surgery can be reduced to a
minimum.
A basic setting is the setting which is defined at the factory or by the user and is stored in the
program memory. Consequently, every program has a basic setting.
The basic setting for program “0“ is defined and programmed by the manufacturer and cannot be
reprogrammed by the user. For the ERBOTOM ICC 350, Software Version 1.05 and up, the basic
setting for program “0“ is defined and stored as follows:
AUTO CUT
:
Auto. voltage control, Effect 3, power limitation 150 watts
Activation with yellow key or yellow pedal
AUTO COAG 1
:
Soft coagulation, power limitation 60 watts
Activation with blue key
AUTO COAG 2:
Forced coagulation, power limitation 60 watts
Activation with blue key or blue pedal
AUTO BIPOLAR
: Power limitation 40 watts
Activation with white pedal, Auto Stop off
The basic settings for programs “1“ to “9“, as well as Program “b“ = Bipolar for bipolar cutting
and coagulation, can be custom-defined and programmed by the user. It is recommended that the
corresponding basic settings be documented.
The use of programs is especially practical for special application areas or cutom settings. The
basic setting of each selected program is also immediately available again if the unit has been
switched off in the meantime. If another program is required than that appearing in display 6.3
after the unit is switched on, the required program can be called up by pressing key 6.2
“PROGRAM“. The unit is immediately ready to operate using this program. If necessary, the
functions and parameters appearing on the front panel for the appropriately selected program
through the basic setting, can be changed or adapted to the immediate purpose and, if the changed
5.4
Auto Stop mode
In the Auto Stop mode, the HF generator is automatically switched off as soon as the
water reaches the vapor phase in the coagulum. The Auto Stop mode is switched once
signal lamp 5.4.1 is illuminated.
5.5
Activation, changeover switch for the blue pedal of a dual-pedal footswitch
If the right signal 5.5.1 is illuminated in this key, the Auto Bipolar can be started via the
blue pedal on a dual-pedal footswitch. Auto Bipolar can also be activated via the white
pedal on a single-pedal footswitch or by Auto Start.
Содержание ERBOTOM ICC 350 M
Страница 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Страница 2: ......
Страница 8: ...1 2...
Страница 10: ...2 2...
Страница 18: ...3 8...
Страница 19: ...7 8 9 10 International Standard...
Страница 20: ......
Страница 42: ...4 22...
Страница 69: ...9 11 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 70: ...9 12 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 71: ...9 13 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 72: ...9 14 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 77: ...9 19 AUTO CUT MODE Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 79: ...9 21 SOFT COAGULATION AUTO BIPOLAR Power output as a function of the power limitation RL 125 ohms...
Страница 80: ...9 22 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms Forced 1 2 3 4...
Страница 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Страница 88: ...11 2...
Страница 94: ...12 6...
Страница 96: ...13 2...
Страница 98: ...14 2...
Страница 100: ...15 2...