10-1
10
INSTALLATION
10.1
Spatial requirements
High-frequency surgery units must only be operated in medically used rooms. The spatial
requirements, in regard to electric installation, affect e.g. the grounded conductor system, the
ground fault interrupt system, as well as measures for preventing electrostatic charges.
The unit is used in rooms in which personnel can pick up electrostatic charges, for example in
rooms with electrically nonconductive floors, thus touching the front panel of the units can lead
to a brief illumination of light diodes or seven-segment displays due to discharge of an electrostatic
charge. However, this occurrence does not change the settings on the front panel.
10.2
Set-up possibilities in the operating room
ICC model series units can be set up in principle on tables, consoles on ceiling suspensions or
wall- mounted arms, as well as on special equipment carts.
10.3
Power connection
High-frequency surgical units must only be connected via the power cord provided by the unit
manufacturer or one of equal quality equipped with the national test symbol to a correctly installed
hospital grade power socket. Here, no outlet power socket or extension cord should be used if
possible for reasons of safety. If their use is unavoidable, these must be equipped with a protective
ground in perfect working order. The power outlet must be secured with a fuse with at least 10 A
rated current.
10.4
Potential equalization
If necessary, the unit can be connected to the potential equalization of the room. This should
prevent low-frequency electrode currents, e.g. low-frequency leakage currents in a defective
protective ground system, from endangering the patient.
ICC model series units are equipped with a potential equalization connector on the unit back
panel according to DIN 42 801. In this way, the units can be connected via a potential equalization
line to a potential equalization terminal at the set-up location.
10.5
Explosion protection
High-frequency surgical units intentionally generate electric sparks between an active electrode
and tissue. Electric sparks may also result within the unit. For this reason, high-frequency surgical
units must not be used in areas where there is a danger of explosion. Considered as explosive is
the area up to 20 cm above the floor and the area around and below the operating table, if
flammable or explosive cleaning agents, disinfectants, anesthetics, etc. are used. High-frequency
surgical units are normally installed outside the zone designated as potentially explosive.
WARNING
Footswitches are used within the potentially explosive area nevertheless and must therefore be
designed as explosion-protected.
Содержание ERBOTOM ICC 350 M
Страница 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Страница 2: ......
Страница 8: ...1 2...
Страница 10: ...2 2...
Страница 18: ...3 8...
Страница 19: ...7 8 9 10 International Standard...
Страница 20: ......
Страница 42: ...4 22...
Страница 69: ...9 11 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 70: ...9 12 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 71: ...9 13 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 72: ...9 14 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 77: ...9 19 AUTO CUT MODE Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 79: ...9 21 SOFT COAGULATION AUTO BIPOLAR Power output as a function of the power limitation RL 125 ohms...
Страница 80: ...9 22 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms Forced 1 2 3 4...
Страница 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Страница 88: ...11 2...
Страница 94: ...12 6...
Страница 96: ...13 2...
Страница 98: ...14 2...
Страница 100: ...15 2...