4-21
You may connect three instruments at the same time to the ICC. For reasons of safety, these can
however only be used alternately. Only one socket ever carries HF voltage.
Each time after switching on the power switch, an automatic test program is started within the
unit which recognizes and signals the following errors in the operating controls for the unit and
for accessories connected to the unit:
1.
If a key on the front panel is shorted or pressed due to an error when the power switch is
switched on, this error is indicated acoustically and by an Error Number after the power
switch has been switched on.
2.
If a key on the electrode handle is shorted or bypassed at low resistance due to an error
(e.g. by moisture in the electrode handle) or pressed while the power switch is switched
on, this error is signaled acoustically and indicated by an Error Number after switching on
the power switch.
3.
If a footswitch contact is shorted due to an error, a pedal sticks or a pedal is pressed while
the power switch is switched on, this error is indicated acoustically and by an Error Number.
CAUTION
Every function field can only then be activated if it has been completely set. If an attempt is
made to activate a function field which has not or not completely been set, the unit produces an
intermittent acoustic warning signal and indicates this operating error by illumination of the
triangle symbols on the corresponding function field.
Содержание ERBOTOM ICC 350 M
Страница 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Страница 2: ......
Страница 8: ...1 2...
Страница 10: ...2 2...
Страница 18: ...3 8...
Страница 19: ...7 8 9 10 International Standard...
Страница 20: ......
Страница 42: ...4 22...
Страница 69: ...9 11 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 70: ...9 12 CUT MODE Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 71: ...9 13 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 300 watts...
Страница 72: ...9 14 HIGH CUT Effect 1 2 3 4 Power output as a function of load resistance for power limitation of 150 watts...
Страница 77: ...9 19 AUTO CUT MODE Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Страница 79: ...9 21 SOFT COAGULATION AUTO BIPOLAR Power output as a function of the power limitation RL 125 ohms...
Страница 80: ...9 22 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms Forced 1 2 3 4...
Страница 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Страница 88: ...11 2...
Страница 94: ...12 6...
Страница 96: ...13 2...
Страница 98: ...14 2...
Страница 100: ...15 2...