Turbimax CUS52D
Technical data
Hauser
43
Response time
> 1 second, adjustable
Detection limit
Detection limit in accordance with ISO 15839 in ultrapure water:
Application
Measuring range
Detection limit
Formazine
0 to 10 FNU (ISO 15839)
0.0015 FNU
13.3 Environment
Ambient temperature
range
–20 to 60 °C (–4 to 140 °F)
Storage temperature
–20 to 70 °C (–4 to 158 °F)
Degree of protection
IP 68 (1.8 m (5.91 ft) water column over 20 days, 1 mol/l KCl)
Electromagnetic
compatibility (EMC)
Interference emission and interference immunity as per
• EN 61326-1:2013
• EN 61326-2-3:2013
• NAMUR NE21: 2012
13.4 Process
Process temperature range
Stainless steel sensor
–20 to 85 °C (–4 to 185 °F)
Plastic sensor
–20 to 60 °C (–4 to 140 °F)
Under high temperatures combined with extremely high or low pH values and chemical
boundary conditions, e.g. during CIP cleaning processes, the sensor has limited long-term
stability.
To avoid damage to the sensor, only use the sensor in combination with a retractable
assembly in CIP cleaning processes. The retractable assembly allows the sensor to be
removed from the process during cleaning.
Process pressure range
Stainless steel sensor
0.5 to 10 bar (7.3 to 145 psi) (abs.)
Plastic sensor
0.5 to 6 bar (7.3 to 87 psi)
Flow limit
Minimum flow
No minimum flow required.
For solids which have a tendency to form deposits, ensure that sufficient mixing is
performed.