Turbimax CUS52D
Operation
Hauser
27
1-point calibration
The measured error between the measured value of the device and the laboratory
measured value is too large. This is corrected by a 1-point calibration.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
x
y
a
b
A0039320
29
Principle of a 1-point calibration
x
Measured value
y
Target sample value
a
Factory calibration
b
Application calibration
1. Select data record.
2. Set the calibration point in the medium and enter the target sample value (laboratory
value).
2-point calibration
Measured value deviations are to be compensated for at 2 different points in an
application (e.g. the maximum and minimum value of the application). This aims to ensure
a maximum level of accuracy between these two extreme values.