Turbimax CUS52D
Diagnostics and troubleshooting
Hauser
35
9
Diagnostics and troubleshooting
9.1
General troubleshooting
When troubleshooting, the entire measuring point must be taken into account:
• Transmitter
• Electrical connections and cables
• Assembly
• Sensor
The possible causes of error in the following table relate primarily to the sensor.
Problem
Testing
Solution
No display, no sensor reaction
• Power supplied to transmitter?
• Sensor connected correctly?
• Buildup on optical windows?
• Connect mains voltage
• Connect sensor correctly
• Clean sensor
Display value too high or too low
• Buildup on optical windows?
• Sensor calibrated?
• Cleaning
• Calibration
Display value fluctuating greatly
Is the mounting location correct?
• Select a different mounting location
• Adjust the measured value filter
Pay attention to the troubleshooting information in the Operating Instructions for the
transmitter. Check the transmitter if necessary.
Содержание Turbimax CUS52D
Страница 2: ......
Страница 46: ......
Страница 47: ......
Страница 48: ...www addresses endress com 71473306 71473306...