Turbimax CUS52D
Operation
Hauser
25
yes
no
start measurement
measurement
correct
?
pipe or assembly
where is the
sensor installed
?
open basin
or channel
measurement
correct
?
activate reflection
compensation
no
yes
calibrate
and
adjust
end
A0030787-EN
28
Flowchart
In order to be able to use the "Offset", "Factor" or "Assembly adjustment" functions, a
new data record must first be generated by means of a 1 to 6-point calibration or by
duplication of a factory data record.
Assembly adjustment
Both the optical design of sensor CUS52D and flow assemblies CUA252 and CUA262 are
optimized to minimize measuring errors caused by wall effects in assemblies or pipes
(measuring error in CUA252 < 0.02 FNU).
You can use the
Assembly adjustment
function to automatically compensate for any
remaining measuring errors caused by wall effects. The functionality is based on formazine
measurements and may thus require a calibration downstream in order to adapt the
measurement to the corresponding application or medium.
Adjustment
Description
PE 100
Adjustment to flow assembly CUA252 (material: polyethylene)
1.4404 (AISI 316 L)
Adjustment to flow assembly CUA262 (material: stainless steel 1.4404)
Customized, standard
Adjustment to any pipe/assembly
Customized, advanced
Adjustment only recommended for Hauser service personnel