Installation
Turbimax CUS52D
16
Hauser
1
2
3
1
4
5
A0030696
12
Example of measuring system with immersion assembly
1
Flexdip CYH112 holder
2
Liquiline CM44x multi-channel transmitter
3
Protective cover
4
Flexdip CYA112 assembly
5
Turbimax CUS52D turbidity sensor
This type of installation is particularly suitable for strong or turbulent flow
> 0.5 m/s (1.6 ft/s) in basins or channels.
5.2.2
Mounting options
Installing with CUA451 retractable assembly
2
A0022285
13
Installing with CUA451 retractable assembly
The installation angle is 90°.
The arrow points in the direction of flow.
The optical windows in the sensor must be
aligned against the direction of flow.
The medium pressure may not exceed
2 bar (29 psi) for manual assembly
retraction.