Turbimax CUS52D
Installation
Hauser
13
Solid state reference
45 (1.77)
58 (2.28)
60 (2.36)
43 (1.69)
86 (3.39)
30
(1.18)
Ø 40 (1.57)
9.5 (0.37)
A0030821
7
Solid state reference Calkit CUS52D. Dimensions: mm (in)
5.1.2
Orientation in pipes
4
2
1
3
5
2
3
4
90°
A0030698
8
Permitted and unacceptable orientations in pipes
• Install the sensor in places with consistent flow conditions.
• The best installation location is in the ascending pipe (item 1). Installation in the
horizontal pipe (item 4) is also possible.
• Do not install the sensor in places where air spaces or bubbles occur (item 3) or where
sedimentation may occur (item 2).
• Avoid installation in the down pipe (item 5).
• Avoid fittings downstream from pressure reduction stages which can lead to outgassing.
Wall effects
Backscattering on the pipe wall may result in the distortion of measured values in the case
of turbidity values < 200 FNU. Therefore a pipeline diameter of at least 100 mm (3.9 in) is
recommended for reflecting materials (e.g. stainless steel). An assembly adjustment onsite
is also recommended.