Installation
Turbimax CUS52D
20
Hauser
Compressed air cleaning
1
2
3
A0030785
22
Installation markings 1 to 3
20°
A0030786
23
Mounting position
Mount the compressed air cleaning system as follows:
1. Fit the compressed air cleaning system on the sensor (→ 23).
2. Position the securing ring for the compressed air cleaning system between
installation marks 2 and 3 (→ 22).
3. Using a 4 mm (0.16 in) Allen key, tighten the securing screw of the compressed air
cleaning system slightly so that the compressed air cleaning system can still be
rotated.
4. Turn the compressed air cleaning system so that the slit on the black ring is on
installation mark 1 (→ 22).
This way the nozzle is offset by 20° when blowing air at the optical windows.
5. Tighten the securing screw.
6. Fit the compressed air hose on the hose connection.
5.3
Post-installation check
Put the sensor into operation only if the following questions can be answered with "yes":
• Are the sensor and cable undamaged?
• Is the orientation correct?
• Has the sensor been installed in the process connection, and does not suspend freely
from the cable?