Principle of a 2-point calibration
x
Measured value
y
Target sample value
a
Factory calibration
b
Application calibration
1. Select a data record.
2. Set 2 different calibration points in the medium and enter the corresponding set
points.
A linear extrapolation is performed outside the calibrated operational range (gray
line).
The calibration curve must be monotonically increasing.
3-point calibration
1
2
3
4
5
6
7
8
1
2
3
4
5
6
x
y
a
b
A0039322
31
Principle of multipoint calibration (3 points)
x
Measured value
y
Target sample value
a
Factory calibration
b
Application calibration