Commissioning
Proservo NMS80
58
Hauser
7.5
Configuration of the inputs
7.5.1
Configuration of the HART inputs
Connecting and addressing HART devices
D
E
F
C
B
A
1
1
1
1
3
2
2
4
1
HR
CDI
WP
on
SIM
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
POWER
i
B
B1-3
B4-8
1 2 3
4 5 6 7 8
C
1 2 3
4 5 6 7 8
C1-3
C4-8
A0032955
29
Possible terminals for HART loops
B
Analog I/O module in slot B (availability depending on device version )
C
Analog I/O module in slot C (availability depending on device version )
E
HART Ex is output (available in all device versions)
HART devices must be configured and given a unique HART address via their own user
interface before they are connected to the Proservo NMS8x
Submenu: Setup → Advanced setup → Input/output → Analog I/O
Parameter
Meaning / Action
Operating mode
Select:
•
HART 4..20mA input
if only one HART device is connected to this loop. In this case the
4-20mA signal can be used in addition to the HART signal.
•
HART master
if up to 6 HART devices are connected to this loop.
3)
The current software does not support HART devices with address 0 (zero).