Commissioning
Proservo NMS80
50
Hauser
7.4
Calibration
After installing or replacing NMS8x or its parts (sensor module, detector unit, wire drum, or
measuring wire), several calibration steps are required. All calibration steps may not be
required, depending on whether the device is being installed, adjusted, or replaced (see table
below).
Type of installation/replacement
Calibration step
Sensor calibration
Reference calibration
Drum calibration
All-in one
Not required
Not required
Not required
Displacer shipped separately
Required
Required
Required
Displacer installation through
calibration window
Required
Required
Required
Replacement/
maintenance
Drum
Required
Required
Required
Displacer
Not required
Required
Required
Sensor module
Not required
Required
Required
Detector unit
Required
Required
Required
7.4.1
Verification of displacer and wire drum
Prior to installation of NMS8x, confirm that all of the following data of the displacer and the
wire drum on the nameplate match with those programmed into the device.
Parameters to be confirmed
Parameters
Navigate to:
Displacer diameter
Setup → Advanced setup → Sensor config → Displacer → Displacer diameter
Displacer weight
Setup → Advanced setup → Sensor config → Displacer → Displacer weight
Displacer volume
Setup → Advanced setup → Sensor config → Displacer → Displacer volume
Displacer balance volume
Setup → Advanced setup → Sensor config → Displacer → Displacer balance volume
Drum circumference
Setup → Advanced setup → Sensor config → Wiredrum
Wire weight
Expert → Sensor → Sensor config → Wiredrum → Wire weight