Commissioning
Proservo NMS80
60
Hauser
7.5.2
Configuration of the 4-20mA inputs
D
E
F
C
B
A
1
1
1
1
3
2
2
4
1
HR
CDI
WP
on
SIM
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
POWER
i
B
B1-3
B4-8
1 2 3
4 5 6 7 8
C
1 2 3
4 5 6 7 8
C1-3
C4-8
A0032464
30
Possible locations of the Analog I/O modules, which can be used as a 4-20mA input. The order
code of the device determines which of these modules is actually present.
Submenu: Setup → Advanced setup → Input/output → Analog I/O
Parameter
Meaning / Action
Operating mode
Select
4..20mA input
or
HART 4..20mA input
Process variable
Select which process variable is transmitted by the connected device.
Analog input 0% value
Define which value of the process variable corresponds to an input current of 4 mA.
Analog input 100% value
Define which value of the process variable corresponds to an input current of 20 mA.
Process value
Check whether the indicated value matches the actual value of the process variable.
1)
There is a
Analog I/O
submenu for each Analog I/O module of the device.