Commissioning
Proservo NMS80
54
Hauser
Calibration procedure
Step
Using displacer
Using offset weight
Description
1.
A0028000
A0028000
• Navigate to: Setup → Calibration → Sensor calibration
→ Sensor calibration
• Input the offset weight for the Offset weight used in
step 3 (0.0 g in case of using the displacer only).
• Input the value for the Span weight used in step 4
(weight of displacer indicated on nameplate).
2.
A0027999
A0028001
• Hold up or remove the displacer.
• Select
for next parameter.
• Measuring zero weight is shown on the display.
• Wait until the Zero calibration shows the Finished
and calibration status shows Idle.
When the displacer is being held up, do not
release it until this step is completed.
3.
A0027999
A0028002
• Confirm that the Offset calibration shows the Place
offset weight.
• Hold up the displacer or attach the offset weight.
• Select
for next parameter.
• Measuring offset weight is shown on the display.
• Wait until the Offset calibration shows the Finished
and Calibration status shows Idle.
When the displacer is being held up, do not
release it until this step is completed.
4.
A0028000
A0028000
• Release the displacer or mount it on the measuring
ring if an offset weight was used in the previous step.
• Select
for next parameter.
• Measuring span weight is shown on the display.
• Confirm that the Span calibration shows the Finished
and Calibration status shows Idle.
• Select the Next.
• Confirm that the Sensor calibration shows the
Finished and Calibration status shows Idle.
This completes sensor calibration procedure.
Do not swing the displacer and keep it in as stable
a position as possible.