Proline t-mass T 150
Electrical connection
Hauser
29
7.2.1
Connecting the cables
E
+
–
1
2
22
23
24
25
26
27
10 (0.4)
3 (0.12)
1
2
22
23
24
25
26
27
24 mm
1.
2.
3.
4.
5.
7.
8.
A0017250
5
Engineering unit mm (in)
‣
NOTICE!
Housing degree of protection voided due to insufficient sealing of the housing.
Screw in the thread without using any lubricant. The threads on the cover are coated
with a dry lubricant.
Reverse the removal procedure to reassemble the transmitter.
For HART communication: When connecting the cable shielding to the ground terminal,
observe the grounding concept of the facility.
7.3
Ensuring the degree of protection
The measuring device fulfills all the requirements for the IP66 and IP67 (Type 4X enclosure)
degree of protection.
To guarantee IP 66 and IP 67 degree of protection (Type 4X enclosure), carry out the
following steps after the electrical connection: