Proline t-mass T 150
Operation options
Hauser
37
8.2.6
User roles and related access authorization
The two user roles "Operator" and "Maintenance" have different write access to the parameters
if the customer defines a user-specific access code. This protects the device configuration via
the local display from unauthorized access .
Access authorization to parameters
User role
Read access
Write access
Without access code
(from the factory)
With access code
Without access code
(from the factory)
With access code
Operator
--
Maintenance
1)
Despite the defined access code, certain parameters can always be modified and thus are excepted from the write
protection, as they do not affect the measurement. Refer to the "Write protection via access code" section
For detailed information about parameters which can always be modified, refer to the
Operating Instructions for the device on the CD-ROM provided
If an incorrect access code is entered, the user obtains the access rights of the "Operator" role.
The user role with which the user is currently logged on is indicated by the
Access status
display
parameter. Navigation path:
Operation
menu →
Access status display
parameter
8.2.7
Disabling write protection via access code
If the -symbol appears on the local display in front of a parameter, the parameter is write-
protected by a user-specific access code and its value cannot be changed at the moment using
the local display .
The locking of the write access via local operation can be disabled by entering the customer-
defined access code via the respective access option.
1. After you press , the input prompt for the access code appears.
2. Enter the access code.
The -symbol in front of the parameters disappears; all previously write-protected
parameters are now re-enabled.
8.2.8
Enabling and disabling the keypad lock
The keypad lock makes it possible to block access to the entire operating menu via local
operation. As a result, it is no longer possible to navigate through the operating menu or
change the values of individual parameters. Users can only read the measured values on the
operational display.
Local operation with mechanical push buttons (display module SD02)
Display module SD02: order characteristic
"Display; Operation"
, option
C