Proline Promass 80S, 83S
Hauser
19
Environment
Ambient temperature range
Sensor, transmitter:
• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Storage temperature
–40 to +80 °C (–40 to +175 °F), pref20 °C (+68 °F)
Degree of protection
Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock resistance
According to IEC/EN 60068-2-31
Vibration resistance
Acceleration up to 1 g, 10 to 150 Hz, following IEC/EN 60068-2-6
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR recommendation NE 21
Process
Fluid temperature range
Sensor
–50 to +150 °C (–58 to +302 °F)
Medium density
0 to 5000 kg/m
(0 to 312 lb/ft
)
Fluid pressure range
(nominal pressure)
Flanges
• According to DIN PN 40 to 63
• According to ASME B16.5 Cl 150, Cl 300
• JIS 10K, 20K, 40K
Secondary containment pressure rating
The sensor housing is filled with dry nitrogen and protects the electronics and mechanics inside.
The following secondary containment pressure rating is only valid for a fully welded sensor housing
and/or a device equipped with closed purge connections (never opened, as delivered).
DN
Secondary containment rating
(designed with a safety factor
≥
4)
Burst pressure of secondary
containment
[mm]
[in]
[bar]
[psi]
[bar]
[psi]
8
³
⁄
₈
"
25
362
190
2755
15
½"
25
362
175
2535
25
1"
25
362
165
2930
40
1½"
16
232
64
925
50
2"
10
145
54
780