Proline Promass 80S, 83S
Hauser
13
Maximum measured error
Design fundamentals
→
15
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = fluid temperature
Base accuracy
Mass flow and volume flow (liquids)
Promass 83S:
• ±0.10% o.r.
Promass 80S:
• ±0.15% o.r.
Mass flow (gases)
±0.50% o.r.
Density (liquids)
• Reference conditions: ±0.0005 g/cm
3
• Field density calibration: ±0.0005 g/cm
3
(valid after field density calibration under process
conditions)
• Standard density calibrations: ±0.01 g/cm
3
(valid over the entire measuring range of the sensor
→
19)
• Special density calibration: ±0.002 g/cm
3
(optional, valid range: +5 to +80 °C (+41 to +176 °F) and
0 to 2.0 g/cm
)
Temperature
±0.5 °C ± 0.005 · T °C (±1 °F ± 0.003 · (T - 32) °F)
Zero point stability
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
Zero point stability
[mm]
[in]
[kg/h]
[lb/min]
8
³
⁄
₈
"
0.20
0.007
15
½"
0.65
0.024
25
1"
1.80
0.066
40
1½"
4.50
0.165
50
2"
7.00
0.257
DN
1:1
1:10
1:20
1:50
1:100
1:500
[mm]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
8
2000
200.0
100.0
40.00
20.00
4.000
15
6500
650.0
625.0
130.0
65.00
13.00
25
18000
1800
900.0
360.0
180.0
36.00
40
45000
4500
2250
900.0
450.0
90.00
50
70000
7000
3500
1400
700.0
140.0