Proline Promass 80S, 83S
Hauser
5
Measuring ranges for gases
The full scale values depend on the density of the gas. Use the formula below to calculate the full scale
values:
max(G)
=
max(F)
·
ρ
(G)
/ x [kg/m
3
]
max(G)
= max. full scale value for gas [kg/h]
max(F)
= max. full scale value for liquid [kg/h]
ρ
(G)
= Gas density in [kg/m
] at operating conditions
Here,
max(G)
can never be greater than
max(F)
Calculation example for gas:
• Sensor type: Promass S, DN 50
• Gas: air with a density of 60.3 kg/m
(at 20 °C and 50 bar)
• Measuring range (liquid): 70000 kg/h
• x = 90 (for Promass S, DN 50)
Max. possible full scale value:
max(G)
=
max(F)
·
ρ
(G)
÷ x [kg/m
] = 70000 kg/h · 60.3 kg/m
÷ 90 kg/m
= 46900 kg/h
Recommended full scale values
See information in the "Limiting flow" section
→
23 ff.
Operable flow range
Greater than 1000: 1. Flow rates above the preset full scale value do not overload the amplifier, i.e. the
totalizer values are registered correctly.
Input signal
Status input (auxiliary input)
U = 3 to 30 V DC, R
i
= 5 k
Ω
, galvanically isolated.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start,
batching start/stop (optional), totalizer reset for batching (optional).
Status input (auxiliary input) with PROFIBUS DP
U = 3 to 30 V DC, R
i
= 3 k
Ω
, galvanically isolated.
Switch level: ±3 to ±30 V DC, independent of polarity.
Configurable for: positive zero return, error message reset, zero point adjustment start, batching start/
stop (optional), totalizer reset for batching (optional).
Status input (auxiliary input) with Modbus RS485
U = 3 to 30 V DC, R
i
= 3 k
Ω
, galvanically isolated.
Switch level: ±3 to ±30 V DC, independent of polarity.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start.
Current input (only Promass 83)
Active/passive selectable, galvanically isolated, resolution: 2
μ
A
• Active: 4 to 20 mA, R
L
< 700
Ω
, U
out
= 24 V DC, short-circuit proof
• Passive: 0/4 to 20 mA, R
i
= 150
Ω
, U
max
= 30 V DC
DN
x
[mm]
[in]
8
³
⁄
₈
"
60
15
½"
80
25
1"
90
40
1 ½"
90
50
2"
90