Proline Promag W 400 HART
Technical data
Hauser
147
16
Technical data
16.1 Application
The measuring device described in this manual is intended only for flow measurement of
liquids with a minimum conductivity of 5 μS/cm.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
sufficiently resistant.
16.2 Function and system design
Measuring principle
Electromagnetic flow measurement on the basis of
Faraday's law of magnetic induction
.
Measuring system
The device consists of a transmitter and a sensor.
Two device versions are available:
• Compact version – transmitter and sensor form a mechanical unit.
• Remote version - transmitter and sensor are mounted in separate locations.
For information on the structure of the device
16.3 Input
Measured variable
Direct measured variables
• Volume flow (proportional to induced voltage)
• Electrical conductivity
In custody transfer: only volume flow
Calculated measured variables
Mass flow
Measuring range
Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
Electrical conductivity: ≥ 5 μS/cm for liquids in general
Flow characteristic values in SI units
Nominal
diameter
Recommended
flow
Factory settings
min./max. full scale value
(v ~ 0.3/10 m/s)
Full scale value current
output
(v ~ 2.5 m/s)
Pulse value
(~ 2 pulse/s)
Low flow cut off
(v ~ 0.04 m/s)
[mm]
[in]
[m
3
/h]
[m
3
/h]
[m
3
]
[m
3
/h]
25
1
9 to 300 dm
3
/min
75 dm
3
/min
0.5 dm
3
1 dm
3
/min
32
–
15 to 500 dm
3
/min
125 dm
3
/min
1 dm
3
2 dm
3
/min
40
1 ½
25 to 700 dm
3
/min
200 dm
3
/min
1.5 dm
3
3 dm
3
/min