Diagnostics and troubleshooting
Proline Promag W 400 HART
124
Hauser
Error
Possible causes
Solution
No or incomplete display of
contents in the Web browser
• JavaScript not enabled
• JavaScript cannot be enabled
1. Enable JavaScript.
2. Enter http://192.168.1.212/
basic.html as the IP address.
Operation with FieldCare or
DeviceCare via CDI-RJ45 service
interface (port 8000)
Firewall of computer or network is
preventing communication
Depending on the settings of the
firewall used on the computer or in
the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
Flashing of firmware with FieldCare
or DeviceCare via CDI-RJ45 service
interface (via port 8000 or TFTP
ports)
Firewall of computer or network is
preventing communication
Depending on the settings of the
firewall used on the computer or in
the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
12.2 Diagnostic information via light emitting diodes
12.2.1 Transmitter
Different LEDs in the transmitter provide information on the device status.
LED
Color
Meaning
Supply voltage
Off
Supply voltage is off or too low
Green
Supply voltage is ok
Link/Activity
Orange
Link available but no activity
Flashing orange
Activity present
Communication
Flashing white
HART communication is active.
Alarm
Green
Measuring device is ok
Flashing green
Measuring device not configured
Off
Firmware error
Red
Main error
Flashing red
Error
Flashing red/green
Start measuring device