Wiring
Proline Promag 55 PROFIBUS PA
48
Hauser
a0005996
Fig. 33:
Connecting the transmitter, cable cross-section: max. 2.5 mm
2
A
View A (field housing)
B
View B (wall-mount housing)
a
Connection compartment cover
b
Cable for power supply: 20 to 260 V AC, 20 to 64 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
c
Ground terminal for protective ground
d
Fieldbus cable:
Terminal No. 26: PA + (with reverse polarity protection)
Terminal No. 27: PA – (with reverse polarity protection)
e
Fieldbus cable shield ground terminal
Observe the following:
– the shielding and grounding of the fieldbus cable
– that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible
f
Service adapter for connecting service interface FXA193 (Fieldcheck, FieldCare)
d
c
e
b
27
25
23
21
2
1
26
24
22
20
L1 (L+)
PA +
N (L-)
PA –
f
PA –
PA +
1
2
c
e
f
b
d
22 23
20 21
24 25 26 27
L1 (L+)
N (L–)
B
a
d
b
a
A
d
b