Proline Promag 55 PROFIBUS PA
Troubleshooting
Hauser
101
9
Troubleshooting
9.1
Troubleshooting instructions
Always start troubleshooting with the checklist below, if faults occur after startup or during
operation. The routine takes you directly to the cause of the problem and the appropriate remedial
measures.
Check the display
No display visible and no
output signals present.
1.
Check the supply voltage
→
Terminals 1, 2
2.
20 to 260 V AC and 20 to 64 V DC: 2 A slow-blow / 250 V
3.
Measuring electronics defective
→
Order spare parts
No display visible, but
output signals are present.
1.
Check whether the ribbon-cable connector of the display module is correctly plugged
into the amplifier board
2.
Display module defective
→
Order spare parts
3.
Measuring electronics defective
→
Order spare parts
Display texts are in a
foreign language.
Switch off power supply. Then press the
OS
keys simultaneously to switch the device back
on. The display text will appear in English (default) and is displayed at maximum contrast.
Measured value indicated,
but no signal at the current
or pulse output
Measuring electronics defective
→
Order spare parts
Æ
Error messages on display
Errors which occur during commissioning or measuring operation are displayed immediately. Error messages consist of a
variety of icons. The meanings of these icons are as follows (example):
– Error type:
S
= system error,
P
= process error
– Error message type:
$
= fault message,
!
= notice message
–
EMPTY PIPE
= Type of error, e.g. measuring tube is only partly filled or completely empty
–
03:00:05
= duration of error occurrence (in hours, minutes and seconds)
–
#401
= error number
"
Caution!
• Also observe the information on
• The measuring system interprets simulations and positive zero return as system errors, but displays them as notice
messages only.
Error number:
No. 001 – 399
No. 501 – 699
System error (device error) has occurred
Error number:
No. 401 - 499
Process error (application error) has occurred
Æ
Incorrect connection to the PROFIBUS master
No connection can be made between the control system and the device.
Check the following points:
Supply voltage
Transmitter
Check the supply voltage
→
Terminals 1/2
Device fuse
Check device fuse
20 to 260 V AC and 20 to 64 V DC: 2 A slow-blow / 250 V
Fieldbus connection
PROFIBUS: check data line
Terminal 26 = PA +
Terminal 27 = PA
−
Fieldbus connector
• Check pin assignment/wiring
• Check connection between connector/fieldbus port.
Is the coupling ring tightened correctly?