Proline Promag 55 PROFIBUS PA
Technical data
Hauser
123
a0002671-ae
Fig. 62:
Promag S remote versions (with PFA or PTFE lining)
T
A
Ambient temperature
T
F
Fluid temperature
HT
High temperature version with insulation
➀
Gray shaded area
→
temperature range from –10 to –40 °C (–14 to –40 °F) applies only to stainless steel flanges
Promag H:
Sensor
• DN 2 to 25: –20 to +150 °C (1/12 to 1": –4 to +300 °F)
• DN 40 to 100: –20 to +150 °C (1½ to 4": –4 to +300 °F)
Seals
• EPDM: –20 to +130 °C (–4 to +266 °F)
• Silicone: –20 to +150 °C (–4 to +300 °F)
• Viton: –20 to +150 °C (–4 to +300 °F)
• Kalrez: –20 to +150 °C (–4 to +300 °F)
Conductivity
Minimum conductivity:
•
≥
5
μ
S/cm for all liquids (incl. demineralized water)
!
Note!
In the remote version, the required minimum conductivity is also influenced by the length of the
cable
Limiting medium pressure
range (nominal pressure)
Promag S:
• EN 1092-1 (DIN 2501): PN 10 (DN 200 to 600/8 to 24"), PN 16 (DN 65 to 600/2½ to 24"),
PN 25 (DN 200 to 600/8 to 24 "), PN 40 (DN 15 to 150/½ to 6 ")
• ANSI B16.5: Class 150 (DN ½ to 24"), Class 300 (DN ½ to 6")
• JIS B2220: 10 K (DN 50 to 300/2 to 12 "), 20 K (DN 15 to 300/½ to 12 ")
• AS 2129: Table E (DN 25/1", DN 50/2")
• AS 4087: Cl. 14 (DN 50/2")
Promag H:
The permissible nominal pressure depends on the process connection and seal:
• 40 bar: flange, weld nipple (with O-ring seal)
• 16 bar: all other process connections
PTFE