OUSAF44
Maintenance
Hauser
25
4.
Hold the lamp socket in the gap between the lamp power supply unit and the reflector
and press it out of its fitting. Do not damage the cable of the circuit board! Then press
the lamp out completely through the reflector.
Dispose of the used lamp in accordance with local regulations for lamps containing
mercury.
5. Check whether the cables of the lamp module are worn, or whether the reflector is
damaged.
Inserting a new lamp
When working with the lamp, use a cloth for lenses or wear talc-free latex gloves.
1.
Clean the new lamp and the reflector with ethanol and a cloth for cleaning the lens.
Do not touch optical surfaces once they have been cleaned!
2.
Guide the new lamp carefully through the reflector and into the socket.
Make sure that the lamp is securely seated in the correct position.